Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
63.0032
O2 sensor no. 1: INFO: Sensor is not
operational yet.
−
Allow sensor to warm up.
63.0080
Temperature measurement for O2:
Voltage is outside specification.
−
see error code 41.0070.
−
Replace the O2 Diaphragm PCB.
63.0081
Temperature measurement for O2:
Temperature is outside specification.
−
see error code 63.0080
63.0090
Measurement difference between the
sensors is too high: O2 measurement
fault:
−
Switch unit off and on again. Calibrate
O2 sensors manually.
−
Replace O2 sensors.
63.0100
Internal error in O2 sensor no. 2 (not
used for the time being).
−
63.0101
Internal error in O2 sensor no. 2 (not
used for the time being).
−
63.0110
Internal error in O2 sensor no. 1 (not
used for the time being).
−
63.0111
Internal error in O2 sensor no. 1 (not
used for the time being).
−
63.0120
O2 measurement faulty, detected by
SISW.
−
see error code 63.0090.