Dr
ä
ger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
•
Mount the new 2 Diaphragm PCB
•
Reassemble the unit (do not mount the O
2
sensors yet).
•
Mount the O
2
sensors.
•
Mount the cover of the O
2
sensors.
•
Perform the safety check and functional check as per the Test Certificate.
Only those contacts are listed for which there is a measurement result. For location of c
see also "Schematics and Diagrams,
Fig.11:
".
After replacing the O2 Diaphragm PCB, it is necessary to recalibrate the offset of the O
measurement electronics (see Service Mode, chapter "
Test Step 22
").
O
2
sensors connection