3.3.2
Equivalent device leakage current test
Connect airway temperature sensor (1) to the
socket of the airway temperature sensor of
Savina and to the tester. Connect test probe
to Savina.
Enter serial number of the airway temperature
sensor:
Subsequent measurements may exceed the
first-measured value by max. 50% but must at
the same time be
≤
1000
µ
A.
Initial value: _______
µ
A
Note:
Always enter the initial value in a new PMS.
I
≤
1000
µ
A
3.3.3
equivalent patient leakage current test
Tester
Savina
L
N
PE
Mains voltage
R
v
mA
Test probe
PS
Equivalent device leakage current test
Equivalent device leakage current test
(1)
P
P
Tester
Savina
Mains voltage
L
N
PE
mA
Test probe
PS
mA
Test probe
equivalent patient leakage current test
equivalent patient leakage current test
(1)