4.2.6
Testing LEDs, 7-segment displays, and LC
display
Call up test step 25
Set ”Test target” value to ”LED”.
Test value:
All LEDs should be evenly lit.
Set ”Test target” value to ”LCD”.
Test value:
The display shows no pixel errors.
Set ”Test target” value to ”7-segment”.
Test value:
All 7-segment displays are lit evenly.
4.2.7
Plausibility test
Compare O2 supply pressure sensor with
ambient pressure sensors.
Test Step 27
: call up this test step.
Cut the O2 supply and relieve the pressure.
Test value:
The three displayed values must not differ by
more than 7%.
Switch off the unit. Cut the connection to the
PC.