![Dräger Savina Service Manual Download Page 18](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836018.webp)
18
A
B
C
1.4.11
"Savina Mobil" trolley
1.4.11.1
Castors
Wheels, of which 2 locking. Check screw
fittings for secure seati ng to 25 Nm.
1.4.11.2
Trolley ac cessories
Check condition and function of the following
accessories (if fitted):
Humidifier holder (optional)
Cylinder holder (optional)
"Multiple socket outlet" retrofit kit (option)
Battery pack (option, in trolley pedestal)
External battery connecting cable
1.4.12
Unit accessories
Check condition and function of the following
accessories (if fitted):
Hinged arm (if fitted)
Temperature sensor (if fitted)
1.4.13
Compressed gas connecting tubes
Compressed gas connecting hoses comply
with national regulations.
For example, compressed gas connecting
hoses with sealing rings, neutral color
for O
2
NIST connectors, 3 m,
for O
2
NIST connectors, 5 m,
1.4.14
Hose systems
Hose systems correspond to specifications in
the
Instructions for Use/Operating
Instructions
.
1.4.15
Adult test lung
Check condition and completeness of adult
test lung.
The adult test lung comprises the following
components:
Face mask angle piece
Catheter connector, ISO, size 7
Breathing bag, 2 L
The breathing bag must not be overstretched.
Z F
Z F
Z F
Z
Z D
Z
Z