Error List Savina Version 2.0
65.0001
Measured value is too high.
−
Carry out Service Mode, test 27.
−
Replace O
2
Valve PCB.
65.0002
Measured value is too low.
−
Carry out Service Mode, test 27.
−
Replace O
2
Valve PCB.
65.0010
Ambient pressure sensor is faulty,
detected by SiSW
−
Carry out Service Mode, test 27.
−
Replace O
2
Valve PCB.
65.0011
A/D converter channel for ambient
pressure sensor is faulty, detected by
SiSW.
−
Carry out Service Mode, test 27.
−
Replace Control PCB.
66.0000
Offset calibration outside range.
−
Carry out Service Mode, test 27.
−
Replace O
2
Valve PCB.
66.0001
Measured value is too high.
−
Carry out Service Mode, test 27.
−
Replace O
2
Valve PCB.