Dr
ä
ger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
1.
After the first breath: Generation of the event "VTi high (1)"
(advisory "! Tidal volume high").
2.
After the third breath: Generation of the event "VTi high (2)"
(warning "!!! Tidal volume high").
3.
Transmission of event to ventilation control to stop inspiration.
There are counters for detection 1 and 2 which lead to reaction 1 or 2.
1.
Detection 1: At the end of a mandatory inspiration, the applied tidal volume is below t
monitoring limit again.
2.
Detection 1: After switching to pure spontaneous breathing mode.
3.
Detection 2: At the end of a spontaneous inspiration, the applied tidal volume is belo
monitoring limit again.
4.
Detection 2: There was no spontaneous inspiration between two mandatory inspira
5.
Detection 2: After switching to pure mandatory mode.