Service Mode Savina Version 2.0
Test step 19 is used to display the measured value of the external A/D converter of the master
processor.
V_FLOW
0016 ... 0049 0.020 V ... 0.060 V (with flow exp. = 0). Flow signal o
expiratory flow sensor S2.
V_FLOW_INSP
0246 ... 0491 0.300 V ... 0.600 V (with flow insp. = 0). Flow signa
inspiratory flow sensor S1.
V_PAW_INSP
0491 ... 0737 0.600 V ... 0.900 V Measurement signal of the insp
pressure sensor S4.2.
V_PAW_EXP
0491 ... 0737 0.600 V ... 0.900 V Measurement signal of the exp
pressure sensor S4.1.
V_FIO2_1
0410 ... 0819 0.500 V ... 1.000 V (if system is free from O2).
Measurement signal of O2 sensor 1.
The measurement electronics are
located on the O2 Diaphragm PCB.
V_REF_O2
2375 ... 2539 2.900 V ... 3.100 V Reference voltage for
V_TEMP_AIR
and
V_TEMP_O2
. The reference
voltage is generated on the O2
Diaphragm PCB from the battery
voltage.