1.4.16
Drug nebulizer (white housing)
1.4.16.1
Condition test
Check the condition of the following
components:
1.4.16.2
Functional test
Fill nebulizer housing with water up to the "3"
mark. Supply dru g nebulizer with 2.0 bar ±
0.2 bar.
Test: Vapor must be produced.
1.4.17
Special accessories (if provided)
Check condition and function of the following
special accessories (if available):
Resutator 2000
Child Resutator 2000
1.5
Unit options and configuration
— Not applicable —
Additional note:
The drug nebulizer, 84 05 000, (black housing), must not be used in Savina.