Dr
ä
ger Med izintechnik GmbH
Repair Information/Error Events Sav ina Version 2.0
5.
Adjust the real-time clock to the real values.
6.
Switch unit off and on again.
7.
Check whether the "BIPAP" and "Autoflow" options are available now.
If the device ID in SW 1.01 has not been generated, the "BIPAP" and "Autoflow" options will b
when updating the software from SW 1.01 to SW 1.10. If this is the case, proceed as follows:
1.
Generate device identification number (device ID).
In order to be able to generate the device ID you need to completely adjust the real-time
clock (tt.mm.jj / hr.min). Then switch the unit off and on again, and read out the device I
from page four of the "Config." menu.
2.
Carry out download using SW 1.10.
3.
Adjust the real-time clock to the real values.
4.
Check whether the "BIPAP" and "Autoflow" options are available now.