![Dräger Savina Service Manual Download Page 16](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836016.webp)
16
A
B
C
1.4
General condition
1.4.1
Accompanying d ocuments
Instructions for use and medical product
logbook available according to operators.
1.4.2
Device rear panel
1.4.2.1
Dust filter set
Replace dust filter, if necessary. Dust filter
must not be pressed together.
1.4.2.2
Ambient air filter
Replace ambient-air filter, if necessary.
1.4.3
Power pack
1.4.3.1
DC connection for external battery.
Connection has no visible damage.
1.4.3.2
Cooling air fan
Switch on the unit. Check condition and
functioning of cooling air fan. Switch off the
unit.
1.4.4
O
2
inlet block.
Inlet block is fitted with NIST, DIN or DISS
connectors.
Check condition and tightness of inlet block.
1.4.5
Inspiration block
Check condition of inspiration block.
1.4.5.1
Inspiratory socket
Check condition of inspiratory socket.
1.4.6
Nebulizer c onnection
The nebulizer connection is marked with a
nebulizer symbol.
1.4.7
Flow sensor mount (flow sensor)
The flow sensor can be inserted into its mount
and moved to the right and to the left without
effort.
1.4.7.1
Flow sensor
The flow sensor is undamaged.
Z
Z
Z
Z
Z
Z F
Z D
Z
Z
Z
Z F
Z