![Dräger Savina Service Manual Download Page 135](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836135.webp)
F
o
r
i
n
t
e
r
n
a
l
u
s
e
o
n
l
y
.
C
o
p
y
r
i
g
h
t
r
e
s
e
r
v
e
d
.
R
5
6
6
4
9
0
0
F
L
1
.
f
m
2
7
.
0
8
.
0
1
A
B
C
Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
41.0060
Operating v15 V of O2
sensors is faulty.
−
Operating v15 V prese
"Repair Procedure
Measuring the
supply voltages
")?
−
Operating v15 V for O2
sensors on O2 Diaphragm PCB
present (test point X4/14(+) and X4/17
(AGND) on the O2 Diaphragm PCB,
see also chapter "Schematics and
Diagrams
O2 DiaphragmPCB
−
Replace the O2 Diaphragm PCB.
−
Carry out SW update
≤
1.10.