Error List Savina Version 2.0
63.0005
O2 sensor no. 2: Compensation with
default temperature.
−
Check device temperature in Service
Mode, test 16, and expose device to
ambient temperature of 5 to 40
necessary.
63.0010
O2 sensor no. 1: Sensor not fitted,
faulty, or not operational yet.
−
Replace O2 sensor no. 1.
−
Install O2 sensor no. 1.
−
Allow O2 sensor no. 1 to stabilize.
63.0011
O2 sensor no. 1: Concentration is
outside specification.
−
Check whether the compressed gas at
the O
2
connection has a concentration
of 100% vol. O2.
−
Check ambient pressure sensors in
Service Mode, test 27.
−
Calibrate O2 sensor no. 1 to 100 vol.%
O2.
−
Replace O2 sensor no. 1.
63.0012
O2 sensor no. 1: Voltage for
temperature measurement is outside
of specification (not used for the time
being).
−