29
A
B
C
4.1.5
On-board electrical system simulation
Short-circuit the two upper sockets on the d.c.
voltage input of Savina using the "
Cable for
charge tester
".
Set the voltage from the d.c. source to 14 V
and wait for the "External battery" LED to light
up green.
Reduce the voltage from the DC source to
11 V.
The LEDs on the front of Savina show the
following operating states:
Mains power supply: green
External battery: green
Internal battery:
Green: Internal rechargeable batteries are
fully charged.
Yellow: Internal rechargeable batteries are
being charged.
Remove the charge tester. Switch off the unit.
4.2
Tests in internal service mode
Switch on Savina and, at the same time, press
and hold the backlighting dim-up/dim-down
key and the "Config" key until the service
mode is activated (see also chapter "Service
Mode
Entering the Internal Service Mode
"
).
4.2.1
Checking the RS232 interface
Test Step 0
: call up this test step.
Connect PC to Savina using "
RS232
extension
" and "
RS232 adapter, crossed
".
Start terminal program and adjust the
following:
COM 1 (or corresponding)
Baud rate: 19200
Data bits: 8
Parity: None
Stop bits: 1
Protocol: None
The settings on the PC must match those in
the Medibus log of the Savina in the
"Configure" menu.
Carry out "Send Logbook" function in internal
service mode.
Test value:
The content of the error log is displayed on
the PC (terminal program).