Alarm Messages Savina Version 2.0
1.
ROM faulty.
1.
A complete ROM test is carried out during the self-test.
2.
A ROM test is also carried out cyclically in normal operating mode during which the R
areas to be tested per program cycle have to be adapted to the run-times.
1.
None.
1.
Generation of event "ROM faulty during self-test" or "ROM faulty during cyclic test" (
"!!! Device failure XX.YYYY").
2.
Triggering of a reset which returns the device to cold start if the error has been detec
during the self-test. "Only" an alarm is generated during cycling testing.