Error List Savina Version 2.0
01.1051
Faulty values from data retention:
Calibration of O2 sensor 2.
−
see 01.0310.
01.2000
Data structure in EEPROM older than
newly installed SW.
−
Disappears after new power-on.
01.2001
Data structure in EEPROM newer than
installed SW.
−
Disappears after new power-on.
−
see 01.0300.
−
Carry out SW download using current
SW.
02.0000
Memory test: Error in the internal RAM.
−
Replace Control PCB.
02.0100
Memory test: Error in the external
RAM.
−
Replace Control PCB.