Instrument Qualification Process
The Plate Type setting in each Gen5 protocol should match the plate you are actually
using.
Synergy LX
_FI_T_SF.prt
Parameter
Default Setting
Plate Type:
Costar 96-well black opaque (#3915)
Detection Method:
Fluorescence intensity
Four Read Steps
Read Step 1: Kinetic
Kinetic loop:
Run Time: 0:00:45
Interval: 0:00:03
Reads: 16
Step Label:
Sensitivity Read
Read Well:
D7
Filter Sets:
Filter set: Green
Gain: Auto, Scale to High Wells, D7, 50000
Read Speed:
Normal
Delay after plate movement: 350 msec
Measurements per data point: 10
Dynamic Range: Standard
Read Height:
8.75 mm
Read Step 2: Kinetic
Kinetic loop:
Run Time: 0:01:35
Interval: 0:00:06
Reads: 16
Step Label:
Sensitivity Read Buffer
Read Wells:
C9..E9
Filter Sets:
Filter set: Green
Gain: Auto, Use first filter set gain from FIRST Read Step
Read Speed:
Normal
Delay after plate movement: 350 msec
Measurements per data point: 10
Dynamic Range: Standard
Read Height:
8.75 mm
Synergy LX User Manual
1501000N Revision A