Error Codes
|
143
Error Code
Description
Troubleshooting
5010
USB port: Serial com
response timeout
Verify that USB device (flash drive or printer),
is attached.
For printers, ensure that cable connector is
firmly plugged in at both ends
For flash drives, ensure that it is firmly seated.
If repeatable or reoccurs, contact Technical
Support.
5011
USB port: Serial com
buffer over run error
5012
USB port: Serial com
command/resp error
5013
USB port: Write file error
5014
USB port: Read file error
5015
USB port: Error creating
directory
5016
USB port: Open file error
5017
USB port: Close file error
5018
USB port: Error changing
directory
5019
USB port: Error writing
to printer
501A
USB port: Error deleting
file
501B
USB port: Error syncing
501C
USB port: No output
destination selected
501D
USB port: No flash drive
detected
7FDD
Take3 plate no longer
supported
In order to create and save a Take3 protocol,
the S/N must be entered and the plate
calibrated.
If this protocol is then used with a different
Take3 plate that has not been calibrated, this
error will result.
To use a new Take3 plate, the serial number
must be entered and the plate calibrated
before selecting it for a protocol.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Summary of Contents for Synergy LX
Page 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Page 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...