Luminescence Test
|
84
Troubleshooting
If the luminescence test fails, try the following suggestions. If it continues to fail, print the
results and contact
Technical Support
.
l
Ensure that the reading is performed through a hole in the filter cube, not through
a glass filter.
l
Verify that the filter cube settings in Gen5 match the filter cube installed in the
reader.
l
If the test continues to fail, the optics block may need to be cleaned. Contact
Technical Support
for instructions.