122
| Appendix
C
:
Error Conditions
General Errors
Error Code
Description
Troubleshooting
0100
Abort error
Assay process aborted
Also known as the Halt command that aborts the running
process without executing the Failure Mode & Effects
Analysis (FMEA) task
The carrier is moved outside, and other axes are moved to
safe positions. The aborted process returns an error status
(0100), indicating that it was aborted.
User aborted protocol or
instrument operation. This
is an informational error.
0300
0302
0303
Memory allocation error
Turn the instrument off and
on. If the error reoccurs,
contact Technical Support.
04XX errors
A/D (analog/digital) data collection failure
A/D converter for measurement circuits has failed or is generating inconsistent or noisy data
For:
• Fluorescence measurements
• Absorbance measurements
• Voltage reference measurements
0401
Fluorescence measurement data converter never
indicated “ready.”
Turning the instrument off
and on may resolve the
problem. If the error
reoccurs, contact Technical
Support.
0411
Fluorescence measurement channel producing
inconsistent or noisy data.
0402
Absorbance measurement data convert never indicated
“ready.”
0412
Absorbance measurement channel producing inconsistent
or noisy data.
040F
Voltage reference measurement data converter never
indicated “ready.”
041F
Voltage reference measurement channel producing
inconsistent or noisy data.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Summary of Contents for Synergy LX