Luminescence Test
|
82
2. On the Harta plate’s Calibration Certificate, locate the NIST measurement for the A2
position and convert it to attomoles:
(A2 NIST measurement*0.02884)
3. Calculate the signal-to-noise ratio:
(A2–Mean of the buffer cells)/(3 * Standard deviation of
buffer cells)
4. Calculate the detection limit:
A2 NIST measurement in attomoles/signal-to-noise ratio
Pass/Fail Criteria
l
If the reader is equipped with the low-noise PMT, the detection limit must be
≤ 75
amol
to pass.
l
If the reader is equipped with the red-shifted PMT, the detection limit must be
≤
500 amol
to pass.
If you do not know which PMT is installed, (#49984 = low-noise PMT; #49721 = red-
shifted PMT, please contact Technical Support.
Gen5 Protocol Reading Parameters
The information in this section represents the recommended reading parameters for the
referenced Gen5 protocol(s).
Synergy LX
_LumTest_Harta.prt
Parameter
Default Setting
Plate Type:
8030015 Harta with 8032028 adapter
Delay Step:
3 minutes
Read Step 1:
Detection Method:
Luminescence
Read Type:
Endpoint
Step Label:
Reference well A2
Read Wells:
A2
Filter Set:
LUM
Gain:
150
Integration Time:
0:10.00 MM:SS.ss
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Summary of Contents for Synergy LX
Page 46: ...22 Chapter 2 Installation Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Page 55: ...Filter Cubes 31 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...
Page 139: ...Sample System Report 115 Synergy LX User Manual 1501000N Revision A BioTek Instruments Inc ...