Error Codes
|
131
Error Code
Description
Troubleshooting
3901
3902
3903
3904
3905
3906
Absorbance reference channel dark range outside limits. Y
= wavelength number in list of wavelengths sent in an
assay protocol.
Humidity is outside
environmental specification
of instrument. Note
specification in user manual
and move to area of lower
humidity.
Too much light in read
chamber. Ensure that:
• The instrument enclosure
is completely installed and
secured.
• The carrier door is closing
fully.
• The filter cube access door
is closed.
3911
3912
3913
3914
3915
3916
Absorbance measurement channel dark range out of
limits. Y = wavelength number in list of wavelengths sent
in an assay protocol.
3AXX errors
Absorbance gain errors
3A10
Absorbance gain out of range while performing a full
range gain calibration.
Possible intermittent flash
failure. Contact Technical
Support.
3A1X
Absorbance gain out of range, where x = optics test
wavelength number.
3EXX errors
Absorbance read saturated. Fail if measurement is ≥ 65535. Either absorbance reference or
measurement channel has saturated.
3E10
Absorbance measurement channel saturation during
carrier auto-calibration when on solid area of jig. The
carrier is moved off the center region and absorbance
integration is performed with the flash on, 8 cycles, and 8
resets. A saturation error is flagged if the measurement
channel signal is higher than 50% of full scale.
Too much light in read
chamber. Ensure that:
• Instrument enclosure is
completely installed and
secured.
• The carrier door is fully
closed.
• The filter cube access door
is closed.
Contact Technical Support.
3E0Y
Absorbance reference channel saturation y = test
wavelength #.
3E1Y
Absorbance measurement channel saturation y= test
wavelength #.
3EX2
Ratio between channels out of range.
3FXX errors
Absorbance signal out of range
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.