System Test
|
89
System Test
System Test
, starting on page 63, describes this test and explains where to find
information on error conditions, as well as sample test reports for Synergy LX.
Test Procedure Using the Touchscreen
1. From the Main Menu, tap
Instrument
, then select the
Configuration
tab.
2. Tap
Start
to begin the system test.
3. When the test finishes, tap
USB Report
to save the test results to a USB flash drive,
to print the test results, or
Exit
to close the screen.
Test Procedure Using Gen5
1. From the Gen5 main screen, select
System > Diagnostics > Run System Test
.
If the test fails during execution, a message box appears in the Gen5. Close the box; the
test report contains the error code that was generated by the failure.
2. When the test is complete, a dialog appears, requesting additional information. Enter
any required information and then click
OK
.
3. The results report appears. it shows either “SYSTEM TEST PASS” or “SYSTEM TEST FAIL
*** ERROR (error code) DETECTED.”
If the test failed, look up the error code in
Appendix B, Error Conditions
, starting
on page 119, to determine its cause. If the cause is something you can fix, turn
off the reader, fix the problem, and then turn the reader back on and retry the
test. If the test continues to fail, or if the cause is not something you can fix,
contact Technical Support.
4. If required, print, sign, and date the report, and store it with your test documentation.
Synergy LX User Manual
1501000N Revision A