144
| Appendix
C
:
Error Conditions
Error Code
Description
Troubleshooting
7FDE
Error determining gain.
Sample signal too low
Verify chemistry is correct for EX wavelength
being used
7FDF
Time and date not set.
Please set them now
Set time and date. If time and date are not
held after unit powered down, battery may
need replacement. See users guide for
instructions.
7FE0
Error parsing assay
response
If repeatable or reoccurs, contact Technical
Support.
7FF2
Error calibrating Take3
plate
7FF3
Error determining bulb
status
7FF4
Wells over-range while
determining AutoScale
Wells are out of range (probably saturated).
The reader will lower gain values for a 2nd try,
but if wells are still out of range, this error will
result
7FF5
AutoScale failed with
over range well