Absorbance Testing
|
66
Peak Absorbance Test
l
Check the filter in the C6 position to ensure it is clean. If needed, clean the filter
with lens paper. Do not remove the filter, and do not use alcohol or other cleaning
agents.
l
Verify that the Peak wavelength information entered for the plate in Gen5
matches the information provided on the test plate’s data sheet.
l
Check the calibration due date on the test plate’s label. If the test plate is overdue
for recalibration, contact BioTek to schedule service.
l
Check the microplate carrier to ensure it is clear of debris.
Alignment Test
l
Ensure that the test plate is properly seated in the microplate carrier.
l
Check the four alignment holes (A1, A12, H1, H12) to ensure they are clear of
debris.
l
Check the microplate carrier to ensure it is clear of debris.
Accuracy Test
l
Check the neutral-density filters to ensure they are clean (positions C1, D4, E2, F5,
G3, H6). If needed, clean the filters with lens paper. Do not remove any filters, and
do not use alcohol or other cleaning agents.
l
Verify that the wavelength/expected OD values entered for the plate in Gen5
match the information provided on the test plate's data sheet.
Repeatability Test
l
Check the neutral-density filters to ensure there is no debris that may have shifted
between readings and caused changes.
l
Check the microplate carrier to ensure it is clear of debris.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.