142
| Appendix
C
:
Error Conditions
Error Code
Description
Troubleshooting
2117
Filter name cannot be
empty
Enter filter name
2D00
Assay definition error
If repeatable or reoccurs, contact Technical
Support.
4000
Stored protocol
checksum verify failed
4010
Stored protocol storage
area full
4020
Protocol names must be
unique
Verify that protocol name has not already been
used. If repeatable or reoccurs, contact
Technical Support.
4030
Protocol names cannot
be empty
Enter protocol name
4040
Blanks defined that are
beyond the limits of the
plate
Ensure that the number of blanks does not
exceed the total number of wells on the plate
4050
No compatible filters
configured
Filters in protocol are not configured in reader.
Verify that correct filters have been installed
and update filter configuration table
Note: 50XX errors indicate problems with a device connected to a USB port, either flash drive or
printer. If the device connected is a flash drive, try a different drive. If a printer, ensure that the
printer is powered on and in the “ready” state.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.
Summary of Contents for Synergy LX