Instrument Qualification Procedures
l
Pipette 150 μL of buffer solution into wells C2–F5. Discard the tips.
l
Pipette 150 μL of the
1 nM SF
solution into column 1.
l
Pipette 150 μL of the 1 nM SF solution into column 2. Mix the wells using the
pipette. Do not discard the tips.
l
Aspirate 150 μL from column 2, and dispense into column 3. Mix the wells using
the pipette. Do not discard the tips.
l
Aspirate 150 μL from column 3, and dispense into column 4. Mix the wells using
the pipette. Do not discard the tips.
l
Aspirate 150 μL from column 4, and dispense into column 5. Mix the wells using
the pipette. Do not discard the tips.
l
Aspirate 150 μL from column 5, and discard the tips.
Alternate/Supplemental Tests Using Methylumbelliferone (MUB)
As an alternative to using Sodium Fluorescein, Methylumbelliferone (“MUB”) can be used
to test the fluorescence system.
Required Materials