Error Codes
|
135
Error Code
Description
Troubleshooting
4700
PMT Fluorescence noise test failure...reference channel
Ensure that instrument case
is completely installed and
secured.
Ensure filter Cube access
door is closed.
If repeatable, contact
Technical Support.
4710
PMT Fluorescence noise test failure...measurement
channel
If repeatable, contact
Technical Support.
48XX errors
Fluorescence offset test error
4800
PMT reference offset (failure if <1607.5 or > 1822.5
counts).
Too much light in chamber.
Ensure that instrument case
is completely installed and
secured.
If repeatable, contact
Technical Support.
4810
PMT measurement offset (saturation), test.
4AXX errors
Fluorescence gain error
4A00
PMT gain out of range
Too much light in chamber.
Ensure that instrument case
is completely installed and
secured.
If repeatable, contact
Technical Support.
4A1Y
A flash fluorescence read error is flagged (4A1y,
y=readset#) if the current PMT sensitivity does not match
the sensitivity specified for the readset.
Perform a system self-test
to reset Gain values.
If repeatable, contact
Technical Support.
4BXX errors
PMT operations or test error
4B0X
PMT Reference channel error
Ensure that door is fully
closed and instrument cover
is secured.
Too much light in chamber.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.