138
| Appendix
C
:
Error Conditions
Error Code
Description
Troubleshooting
5002
5202
Probe height Z failed to home
Shipping bracket not
removed.
An object may be
obstructing the path.
If repeatable, contact
Technical Support.
5003
5203
Absorbance Order sorting filter wheel failed to home
If repeatable, contact
Technical Support
5004
5204
Monochromator grating failed to home
51XX and 53XX errors
Autocalibration
5100
5300
Carrier X failed autocalibration
If repeatable, call Technical
Support
5101
5301
Carrier Y failed autocalibration
5104
5304
Absorbance monochromator failed autocalibration
540X errors
Verify movements
5400
Carrier X failed positional verify
Verify there are no
obstructions and that the
carrier and the
dispenser/probe Z-Axis are
not hitting anything. If
repeatable, contact
Technical Support.
5401
Carrier Y failed positional verify
5402
Probe Z-Axis failed positional verify
5404
Monochromator failed positional verify
If repeatable, contact
Technical Support.
55XX errors
5500
Carrier X-Axis not homed successfully
Run self-test or power cycle
instrument If repeatable or
reoccurs, contact Technical
Support.
5501
Carrier Y-Axis not homed successfully
5502
Vertical Z-Axis not homed successfully
5504
Monochromator not homed successfully
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.