Error Conditions
Error Code
Description
Troubleshooting
4B1X
PMT Measurement channel error
Ensure that instrument case
is completely installed and
secured.
If repeatable, contact
Technical Support.
4EXX errors
PMT saturation errors
4E0X
PMT reference channel saturation
Too much light in chamber.
Ensure that instrument case
is completely installed and
secured.
Ensure that instrument door
is fully closed when carrier
enters chamber.
Sensitivity set to high: try
adjusting auto-sensitivity
setting.
Door not closing completely.
If repeatable, contact
Technical Support.
4E1X
PMT measurement channel saturation
4FXX errors
Fluorescence signal out of range
4F00
Fluorescence reference signal out of range
Sensitivity too high.
Ensure that door is fully
closed.
Ensure instrument cover is
installed and secured.
Open hole in emission
and/or excitation filter
cubes.
Ensure that Gen5
Fluorescence/Luminescence
wavelengths table matches
the filters physically installed
in the filter cubes.
Synergy LX User Manual
1501000N Revision A
BioTek Instruments, Inc.