Glucose Control Testing
55
Information stored during
glucose control testing
The following information is stored for every glucose
con trol test using control solution:
n
Glucose control test result
n
Lot number of the control solution
n
Operator ID (if configured)
n
Level of control solution (L1 or L2)
n
Lot number of the test strips
n
Time and date of test
n
Comments (if applicable)
n
Out of range measurements
Control solutions
For blood glucose test strips, control solutions have two
levels:
n
Level 1 (L1): Lo (low values in test results)
n
Level 2 (L2): Hi (high values in test results)
Preparing to run a glucose
control test
Aside from special preparations (see the following
sec tion), a glucose control test is run in the same manner
as a patient test:
n
At least one code file for test strips must be stored
in the meter and match the lot number of the test
strips used (see Chapter 6).
n
The proper test strips must be available.
n
An operator ID must be entered (with password, if
applicable), if the meter is configured for login.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......