Maintenance and Care
134
Deleting patient data
This feature allows you to delete all patient-related data
(patient lists and results). This may become necessary to
comply with data protection and privacy policies when
sending a meter back to the manufacturer or passing it
on to another facility.
Setup Menu
Date / Time
QC
Test Display
Password
Test Ranges
Reagents
Operator ID
Patient ID
Unlock
Delete Pat.
12:48 pm
09/18/13
Test Display
Test Ranges
Setup Menu
Date / Time
Reagents
F
G
H
I
J
K
L
M
N
O
P-Z
123
D-201: Delete Data:
Are you sure you want to
delete all patients and all
results from this meter?
This operation cannot be
undone.
Patient Confi rmation
12:48 pm
09/18/13
To ensure deleted data is not transferred back through
an unwanted automatic synchronization with the DMS,
remove the battery pack immediately after deleting
patient data.
Be aware that the function
Delete Patient
irrevocably
deletes all data once you touch
to confirm. The
system will delete any patient results regardless of
whether they have been synchronized with a data
management system or not. Therefore, always
ensure that you have downloaded all relevant patient
results before you use this function.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......