Patient Glucose Testing
48
Results screen
The hourglass icon indicates the test is running. When the
test is completed and the result is ready, the meter beeps
again.
Patient Test
Patient 123456789
Date 09/18/13 12:48 pm
150
L22
12:48 pm
09/18/13
Range
I-502: Normal Range:
Critical Range:
Reportable Range:
70-200 mg/dL
40-301 mg/dL
11-600 mg/dL
Patient Test
Patient 123456789
Strip Lot 545794
12:48 pm
09/18/13
Patient Test
Range
150
Patient 123456789
Date 09/18/13 12:48 pm
mg/dL
09/18/13
12:48 pm
When the result is displayed, a message or warning may
also appear (depending on system configuration)
notify ing you if the result exceeds the specified limit
values. Additionally, a
red flashing arrow
(see
Patient
Test
screen below) next to the test result indicates that a
result is out of range.
: the value is above the normal/critical upper range.
: the value is below the normal/critical lower range.
These limit values define ranges that can either be
con figured individually by the system administrator in line
with facility guidelines, or are the (technical) limits of the
system. The characteristics of these ranges are explained
on the following page.
The results screen contains a button that changes its
name based on the result (
Range
or
Out of ... Range
).
Touch this button to display the configured limit values.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......