Introduction
22
Roche supports industry wireless standards and
recom mends using products that have Wi-Fi certification.
This certification tests products to the 802.11 industry
stan dards for basic connectivity, security, authentication,
Quality of Service (QoS), interoperability and reliability.
The Wi-Fi CERTIFIED
TM
logo is an assurance that the Wi-Fi
Alliance
®
has tested a product in numerous
configurations and with a diverse sampling of other
devices to ensure compatibility with other Wi-Fi
CERTIFIED equipment that operates in the same
frequency band. The Wi-Fi Alliance network of
independent test labs conducts interoperability testing
programs to ensure that wireless devices work together
and support secure connections.
The ACCU-CHEK Inform II system complies with FCC
radiation exposure limits set forth for an uncontrolled
environment. This equipment should be installed and
operated with mini mum distance of 8 inches (20 cm)
between the radiator and your body.
This transmitter must not be co-located or operated in
conjunction with any other antenna or transmitter.
Changes or modifications made to this equipment not
expressly approved by Roche Diagnostics may void the
FCC authorization to operate this equipment.
This device complies with Part 15 of the FCC Rules and
with RSS-210 of Industry Canada. Operation is subject to
the following two conditions:
(1) this device may not cause harmful interference,
and
(2) this device must accept any interference received,
including interference that may cause undesired
opera tion.
The ACCU-CHEK Inform II system complies with the
emission and immunity requirements described in
IEC 61326-2-6:2006. It has been designed and tested to
CISPR 11 Class B.
In a domestic environment it may cause radio
interference, in which case, you may need to take
measures to mitigate the interference.
The electromagnetic environment in which the ACCU-
CHEK Inform II device will be operated should be
thoroughly evaluated prior to operation of the device.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......