Patient Glucose Testing
41
You now have three different options, depending on
setup, for assigning the subsequent test to a patient.
The patient ID function can be configured by your
system administrator to:
n
Enter any combination of up to 20 alphanumeric
characters, with specified minimum and maximum
lengths.
Alphanumeric characters are any combination of
A - Z and 0 - 9, additionally “.” (period), or “-” (hyphen)
may be used.
n
Enter patient ID via barcode scanner.
n
Select a patient from a list. *
The following options are available for validating
patient IDs:
n
Manual entries can be validated based on an
downloaded list.
n
Confirmation of patient name, date of birth, and ID
can be required. **
n
Barcode masking can be used to eliminate any
characters not belonging to the Patient ID. See
information on “Operator and patient ID barcode
masks” on page 158.
* A list of supported barcode symbologies can be found in
Appendix A2.
** Depending on the DMS you are using, this feature may
be disabled.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......