Initial Startup
116
Value ranges (normal, critical, reportable)
This menu allows you to set limit values for results as
Normal
,
Critical,
or
Reportable
. Results outside these
limits will be flagged to alert you of this event. The
Reportable Range
allows the system administrator to set
an institution defined range for reporting patient results.
1
2
3
4
5
6
7
8
9
0
Report. Range Lo
11
mg/dL
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Normal Range Lo
70
mg/dL
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Critical Range Lo
40
mg/dL
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Report. Range Hi
600
mg/dL
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Normal Range Hi
200
mg/dL
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Critical Range Hi
301
mg/dL
12:48 pm
09/18/13
Values for range limits are entered one directly after the
other in the following order:
n
Normal (low/high limit value)
n
Critical (low/high limit value)
n
Reportable (low/high limit value)
1
Use the keypad to enter the first value.
2
Touch
to enter the following value, or
3
Touch
to return to the previous view.
4
After entering the last value, touch
to save all
the previous settings.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......