Initial Startup
115
Options for glucose control tests
This menu contains options that allow you to specify
whether and at what intervals glucose control tests are
performed. If you require control tests, check your
set tings for STAT tests (see page 112).
1
Touch the desired option to enable it:
n
Control Frequency
–
Always OK
: QC Lockout is disabled, patient tests
can be performed at any time. This applies
regardless of the result of running glucose control
tests.
–
Last QC OK
: QC Lockout is activated only when the
last glucose control test is out of range or a control
test is not available (e.g., when changing the test
strip lot).
–
Hours
: QC Lockout is activated after the specified
time. When a glucose control test is run
successfully, the counter is reset to “0”.
–
Strip Count
: QC Lockout is activated after reaching
the specified number of tests. When a glucose
control test is run successfully, the test strip
counter is reset to “0”.
–
Shift
: QC Lockout is activated per shift.
–
Time Of Day:
QC Lockout is activated at the
specified time of day. (Up to six times per day can
be configured.)
–
Rotating QC:
QC tests require alternating use of L1
and L2 controls, instead of requiring both controls
when performing a QC test.
2
Touch
to save the settings. If the
Hours
or
Strip
Count
option is enabled, enter the desired number,
or:
3
Touch
to exit this menu without saving any
changes.
The display returns in both cases to the
Setup Menu
.
QC Frequency
Control Frequency
Last QC OK
Always OK
Strip Count
Hours
Rotating QC
Shift
Time Of Day
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
QC Hours
50
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
QC Strips
50
12:48 pm
09/18/13
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......