Linearity Testing
87
In the
Linearity Test
menu, the levels available for the
lin earity test are displayed.
4
Touch
L1
to start the subsequent test with this
(first) level.
Inserting test strips
After selecting the level, a flashing green arrow appears
on screen and prompts you to insert the test strip.
Linearity Test
Linearity 12345678 L1
Strip Lot 545794
12:48 pm
09/18/13
1
Remove the test strip from the test strip vial and
close the vial again with the cap.
2
Hold the test strip so the lettering “ACCU-CHEK” is
facing upward.
3
Slide the test strip into the test strip port as far as it
will go in the direction indicated by the arrows on
the test strip.
The meter beeps. The hourglass icon appears and
indicates that the meter is checking the test strip.
Do not apply linearity solution while it is displayed.
Linearity Test
L1
L5
L3
L6
L4
L2
Linearity 12345678
Strip Lot 545794
12:48 pm
09/18/13
Linearity Test
Linearity 12345678 L1
Strip Lot 545794
12:48 pm
09/18/13
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......