USER’S 

MANUAL

In-Vitro Diagnostic

aid for self testing

of GLUCOSE

Summary of Contents for OGCare

Page 1: ...USER S MANUAL In Vitro Diagnostic aid for self testing of GLUCOSE...

Page 2: ...fessional Information OGCARE Meter OGCARE Test strips Explanation of the display Disinfection before test Performing a Blood Glucose Test Flagging Test Results Handling the Used Test Strip Alternative...

Page 3: ...Test Results Performing the Check Strip Test Cleaning your meter Replacing Battery Screen Messages Troubleshooting Specifications Warranty Symbols EMC Test results Disposing of your meter Warrant Card...

Page 4: ...uantitative measurement of the concentration of glucose in whole blood from fingertip palm and forearm by diabetic patients or healthcare professionals as an aid in the management of diabetes It is in...

Page 5: ...This current is measured by the OGCare Meter and displayed as your blood glucose result The strength of these currents changes with the amount of glucose in the blood sample The OGCare Meter automati...

Page 6: ...affecting the system s accuracy Don t decompose demolish repair or transform the system at any time Don t put the system near any electromagnetic field e g TV microwave oven mobile phone Don t allow t...

Page 7: ...ults Never reuse a test strip that has had either blood or control solution applied to it The OGCareTest Strip is for one time use only Before the test you must make sure that the code number on the m...

Page 8: ...minutes before testing The OGCare Blood Glucose Monitoring System shows accurate results between 10 40 C 50 104 F Do not store the meter test strips lancing device and lancet in a place where the chil...

Page 9: ...facility When using the OGCare meter follow always the recognized procedures for handling objects that are potentially contaminated with human material Practice the health and safety policy of your l...

Page 10: ...10 OGCARE METER Display M Button FRONT Test strip slot C Button...

Page 11: ...11 USB port BACK Battery lid Serial number...

Page 12: ...th the OGCare Meter Please check the expiration date of test the strips before use Please write the opening date of the new test strip vial in order to avoid using the expired products Lot Number EXP...

Page 13: ...ow Battery Warning Stored test result Test results Unit of measurement Temperature error No beep Pre post meal mark Ready to test Control solution Average result Month Date Time 1 2 1 2 3 4 5 6 7 8 9...

Page 14: ...clean hands with soap and water and then dry then well You can also use an alcoholic cotton cloth to clean the finger Please wait until the alcohol disappears completely DISINFECTION BEFORE TEST TESTI...

Page 15: ...se to remove it 2 Insert the lancet into the lancet holder and push down firmly until it is fully seated Do not twist the lancet 3 Twist the protective disk until it separates from the lancet Replace...

Page 16: ...er identifies the code number automatically Compare the code number display with the code number on the test strip vial If they do not match try again with an another test strip If the problem happens...

Page 17: ...ck the barrel until it clicks If it does not click the lancing device may have been activated when the lancet was inserted 8 Hold the lancing device firmly against your finger Press the release button...

Page 18: ...hear the beep sound you have enough blood in the test strip After five seconds the test result will appear NOTICE When the blood sample is not sufficient the display will show the message of Er 4 Ple...

Page 19: ...UTION Do not drop the blood directly on the end of the test strip The sample channel at the end of the strip should be filled with blood When you hear the beep sound you have enough blood in the test...

Page 20: ...the beep sound the screen will display test result After taking the strip out the meter will be turned off automatically 11 12 Display for unit mg dL Display for unit mmol L NOTICE If the test result...

Page 21: ...n the device after use and discard it properly If the lancet cap has loosened or fallen please do not use it Dispose always the used lancet in a biohazard container Please store the lancing device in...

Page 22: ...mark using M or C button with the test result shown on the display and the test strip still in the meter When you push the M or C button the meal mark or control solution mark will be toggled NOTICE T...

Page 23: ...ime When you ask the doctor please take the logbook out it is helpful for your treatment 2 Use the strip ejector located backward of the meter to remove the used test strip 3 After taking the test str...

Page 24: ...ternative site testing only two hours or more after taking insulin two hours or more after a meal two hours or more after exercise 3 Do not use alternative site testing if you are pregna nt or if you...

Page 25: ...he site results do not agree with the way you feel after a meal after exercise during illness during times of stress CAUTION Do not change your treatment because of just one result Never ignore sympto...

Page 26: ...your forearm or palm is different from getting a sample from your fingertips Forearm sampling Choose a fleshy area of the forearm away from bone visible veins and hair Sometimes there is less blood f...

Page 27: ...ch may cause your blood sample to smear NOTICE You may need to wait a little longer to get a large drop of blood from the forearm or palm Do not squeeze the site excessively If the sample drop of bloo...

Page 28: ...ton for 3 seconds you will hear the beep the screen will show the whole symbols of the system 2 After 1 second the SEt will be displayed on the screen USING THE SET UP MODE 2 1 METER SETUP MEMORY AND...

Page 29: ...4 Setting Date when date blinks press and release C button until date appears Press M button to set and move to time set mode 5 Setting Time Format the meter can be set in an AM PM 12 hour default fo...

Page 30: ...ting Minute when minute blinks press and release C button until correct minute appears Press M button to set and move to year set mode 8 Setting Year entering the year mode the screen will show the nu...

Page 31: ...Otherwise please contact your Customer Service 9 9 CAUTION Use of the wrong glucose measurement unit may cause you to misinterpret your blood glucose level and incorrect treatment Consult your healthc...

Page 32: ...OFF your new meter has the beeper preset to On You can set the beeper to Off with pressing C button if you prefer this will not work during the system functioning Press M button to set and move to the...

Page 33: ...RESULTS 1 Turn on the meter by pressing button M for 3 seconds to enter the Set Up mode 2 Display shows the SEt when enter Set Up mode 3 Press and release M button until the screen will show the word...

Page 34: ...he option with C button Press M button to move to the next set mode 5 If you wish to delete the all test results from the memory toggle the option to set YES and press M button 6 After deleting the sc...

Page 35: ...button you will hear a beep sound and the meter will be turned on 2 The screen will display the date and time After 2 seconds the last 14 days average result will be displayed Display per mg dL Displ...

Page 36: ...the screen will show the previous test result 5 After completing the reviewing press M button to turn off the meter NOTICE If you don t make any changes the meter will be turned off automatically aft...

Page 37: ...the OGCare Meter to a computer where they can be summarized in a report with graphs and tables To make use of this feature you need OGCare Link Software and the OGCare Link Cable For more information...

Page 38: ...d perform a control test when You open a new box of test strips You left the test strip container open You want to check the meter and test strips Your test strips were stored in extreme conditions of...

Page 39: ...tstrip codes match If they do not please contact your Customer Service 2 Shake well before using the control solution Discard the first drop Wipe the dispenser tip Hold the drop to the narrow channel...

Page 40: ...Marked control solution tests will not be included in your averages Display per mg dL Display for mg dL Display for mmol L 4 CAUTION If you continue to get control solution test results that fall out...

Page 41: ...8 1 Example The label on the vial of test strips show acceptable range for both level 1 and 2 of control solution The result obtained should fall within this range Be sure to compare the result with t...

Page 42: ...st the entire system please rely on the results from testing a control solution 1 Begin with the meter off insert the OGCare Glucose Check Strip in the meter s test strip slot 2 If the OK Message is n...

Page 43: ...ened with one of these cleaning solutions 70 isopropyl alcohol Mild dishwashing liquid mixed with water 10 household bleach solution made on the same day Make sure you squeeze off excess liquid from t...

Page 44: ...indicates that the energy left could operate testing about 50 times more In this period we suggest that you should replace the battery After replacing the battery the stored results will remain but t...

Page 45: ...t strip after the blood symbol appears on the screen Temperature is too hot or too cold for the system to work properly The operating temperature should be 10 40 C 50 104 F Please use the system after...

Page 46: ...he result is higher than 600mg dL 33 3mmol L It indicates that the value of your blood glucose is too high Retest with a new test strip If Hi mark appears again please call your healthcare professiona...

Page 47: ...the test strip is used Check whether the test strip has become invalid Check whether the test strip vial opening date has passed more than 3 months Make sure the meter code and test strip code match T...

Page 48: ...ISPLAY OPERATING TEMP OPERATING HUMIDITY STORAGE TEMP DIMENTIONS WEIGHT OGCARE GLUCOMETER PLASMA EQUIVALENT 0 5 L OGCARE STRIPS 20 600 MG DL 1 1 33 3 MMOL L 5 SECONDS DC 3V CR2032 X 2 APPROXIMATELY MO...

Page 49: ...the following conditions The meter is damaged as a result of incorrect use The meter is damaged as a result of natural disasters e g typhoon earthquake etc The meter is damaged as a result of demoliti...

Page 50: ...ause erroneous results Do not use this instrument in proximity of sources of strong electromagnetic radiation as these may interfere with the proper operation EMC TEST RESULTS DISPOSING OF YOUR METER...

Page 51: ...presentative pharmacies or your medical surgical supply dealer Control Solutions Glucose Control Solution Level 1 Glucose Control Solution Level 2 Check Strip Connecting cable Name Age Sex M F Address...

Page 52: ...BIOCHEMICAL SYSTEMS INTERNATIONAL S R L Loc Palazzo del Pero 23 Arezzo AR 52100 ITALY tel 39 0575 984164 biosys biosys it www biosys it MOGC ED2016 APRILE ENG...