Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
69
5
Place the code key reader on a level surface such
as a bench. Hold the meter 4-6 inches above the
code key reader so that a connection can be made
between the two infrared windows.
6
Touch
to begin downloading data.
n
The code file is ready for transmission as long as
the LED on the code key reader is flashing green,
even if the code key is removed.
–
After the LED stops flashing, remove the code
key and insert a new code key for download, if
required.
– If you get an error message that the download
was not successful, re-insert the same code key
and try again.
n
Do not exchange code keys while the code key
reader is still flashing. If the code key reader is still
flashing, it will continue to transmit the previously
loaded code file and ignore the code file on the
newly inserted code key. You may get an error
message on the meter.
Once the connection is made, the meter provides you
with status information on the download.
Add Strip Lot
Please Wait
I-302: Connecting to code
key reader...
12:48 pm
09/18/13
Place Meter above the Code
Key Reader and then press
the Right Arrow button to
start.
Add Strip Lot
12:48 pm
09/18/13
Add Strip Lot
Please Wait
I-303: Receiving code key
contents...
12:48 pm
09/18/13
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......