Glucose Control Testing
61
Applying the control solution
Once the meter has detected the test strip, the hourglass
icon disappears and you are prompted to apply control
solution.
Control Test
Control L2 (Hi) 123456
Strip Lot 545794
12:48 pm
09/18/13
1
Wait until the flashing drop appears in the display
before applying the control solution. The meter
beeps.
2
Apply a drop of glucose control solution to the
front edge
of the test strip. Do
not
apply the
control solution to the top of the strip.
The control solution is pulled into the test strip by
capillary action.
Once sufficient control solution has been detected, the
meter beeps and the measurement begins.