Appendix
151
Subject/Attribute
Range
Default
Device
DMS
Normal range HI limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y
Y
Normal range LO limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y
Y
STAT tests allowed
0: no
1: yes
1
Y
Y
Number of STAT tests allowed
0 – 9
9
Y
Y
Reportable range HI limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y
Y
Reportable range LO limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y
Y
Reportable range message
Whether to display a
reportable range
message (1) or not (0)
1
N
Y
Reportable range message text
0 – 100 characters
“”
N
Y
Strip limitations warning
Configuration, whether to
display the strip
limitations warning (1) or
not (0).
0
Y
Y
Strip lot verification
0: display only
1: yes/no confirmation
2: list selection
3: scan only
1
Y
Y
Linearity lot verification
0: display only
1: yes/no confirmation
2: prompt for lot entry
1
Y
Y
Display
Contrast
1 - 15
7
Y
Y
Formats and Language
Date format
1: MM/DD/YY
2: DD.MM.YY
1
Y
Y
Time format
1: 24 hours
2: 12 hours
2
Y
Y
Language setting *
1: German
3: French
4: Spanish
5: Italian
6: Dutch
7: Swedish
8: English (USA)
9: Danish
11: Portuguese
8
Y
Y
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......