Troubleshooting
141
Meter reset
A meter reset should only be performed if all other
remedies have failed.
1
Place the meter on a level surface.
2
Press the On/Off button for at least 12 seconds and
release the button.
n
The meter powers off and on again.
n
Roche logo is displayed.
If the Roche logo does not appear within 60
seconds, place the meter on a base unit for a
minimum of 15 minutes to recharge the battery.
n
The meter performs a system check. During the
system check, the
Versions
screen appears for a
few seconds, then disappears. The
Versions
screen
shows the version numbers of the meter
components. (The
Versions
screen shown here is
for illustration purposes only. Version numbers on
your meter may differ).
n
A message appears informing you that the date
and time settings have been lost due to the meter
reset.
3
Confirm the message
Loss of Date/Time
.
SP
.HUQHO
5RRW)6
$SS
%RRW
%XLOG
9HUVLRQV
Checking system...
Performing self-checks...
Performing self-checks...
W-1302: D/T settings lost
due to meter reset or loss of
power. Enter D/T manually,
then synchronize meter or
see System Administrator
before performing further
tests.
Loss of Date/Time
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......