Maintenance and Care
125
Acceptable
Cleaning /
Disinfecting
Agents
To purchase approved disinfecting products, please refer to your facility’s
guidelines, which may include instructions on direct purchase from the
manufacturer or through an approved distributor. For additional
information on product availability for purchase, refer to manufacturer’s
website: http://www.cloroxprofessional.com for Clorox
®
Germicidal Wipes
or http://www.pdipdi.com/ for Super Sani-Cloth
®
Germicidal Disposable
Wipes.
n
Dispose of acceptable cleaning and disinfection solutions per facility
guidelines.
Acceptable active ingredients and products for cleaning and disinfecting are:
–
Clorox
®
Germicidal Wipes
(EPA* reg. no. 67619-12)
Pre-moistened disinfecting cloths (active ingredient 1% (or less) solution
of sodium hypochlorite in water)
–
Super Sani-Cloth
®
Germicidal Disposable Wipes
(EPA* reg. no. 9480-4)
Pre-moistened disinfecting cloths (active ingredient 0.5% quaternary
ammonium chlorides and up to 60% isopropanol)
n
Cleaners and disinfectants were validated separately and only one
cleaning/disinfecting solution should be used on the device as the effect
of using more than one cleaner/disinfectant interchangeably has not been
evaluated.
NOTE:
Always use Clorox Germicidal Wipes (EPA* reg. no. 67619-12)
or Super Sani-Cloth Germicidal Disposable Wipes (EPA* reg. no.
9480-4) to clean and disinfect the meter. Do not use any other
cleaning or disinfecting solution. Using solutions other than
Clorox Germicidal Wipes (EPA* reg. no. 67619-12) or Super
Sani-Cloth Germicidal Disposable Wipes (EPA* reg. no. 9480-4)
could result in damage to the system components.
NOTE:
Roche has tested the approved cleaning and disinfecting product
for a total of 12,045 cycles, which is equal to cleaning and
disinfecting after each patient use up to 11 tests per day over the
3 year use life of the device.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......