Applying a proficiency sample
Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply the
proficiency sample.
Proficiency Test
Sample 123456789
Strip Lot 545794
12:48 pm
09/18/13
1
Wait until the flashing drop appears in the display
before applying the sample. The meter beeps.
2
Apply a drop of sample to the
front edge
of the
test strip. Do
not
apply the sample to the top of the
test strip.
The sample is pulled into the test strip area by
capillary action.
Once sufficient sample has been detected, the meter
beeps and the measurement begins.