Linearity Testing
88
Applying a linearity test sample
Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply linearity
solution.
Linearity Test
Linearity 12345678 L1
Strip Lot 545794
12:48 pm
09/18/13
1
Wait until the flashing drop appears in the display
before applying the solution. The meter beeps.
2
Apply a drop of the linearity solution to the
front
edge
of the test strip. Do
not
apply the solution to
the top of the strip.
The linearity solution is pulled into the test strip
area by capillary action.
Once sufficient linearity solution has been detected, the
meter beeps and the measurement begins.