![background image](http://html.mh-extra.com/html/accu-chek/inform-2/inform-2_operators-manual_455938145.webp)
General Product Information
144
Specification
Meter
Base unit
Power supply unit
Battery voltage/type
3.7 volt rechargeable
(lithium technology)
N/A
N/A
Input voltage
+7.5 V DC
+7.5 V DC
100 to 240 V AC
Input frequency
DC
DC
50 to 60 Hz
Input current
1.7 A (max)
1.7 A (max)
350 to 150 mA
Battery capacity
50 (subsequent)
measurements possible
after 120 min of charging
N/A
N/A
Interfaces
Charge contacts
IR port
Barcode scanner
WLAN
(Channel 1-11 only)
Charge contacts
IR port
RJ45 Ethernet *
USB type B *
DC connector
Replaceable
AC input contacts
Data transfer rate
WLAN: up to 54 Mbps
IR: 9.6K - 115K bps *
Ethernet: 10 Mbps *
USB: 12 Mbps *
N/A
Supported barcodes
Code 128, Code 39,
Code 93, EAN 13**,
Interleaved 2 of 5
+
,
Codabar., GS1 DataBar
Limited, QR Code,
DataMatrix, PDF417
N/A
N/A
*Not applicable to ACCU-CHEK Inform II Base Unit Light
**EAN 13 barcodes are not recommended for identification of operators and patients.
+
Interleaved 2 of 5 codes can only be decoded if they contain a check digit.
Specification
Accessory box
Code key reader
Base unit HUB
Height
3.35 in / 85 mm
0.72 in / 18.4 mm
1.38 in / 35 mm
Width
11.02 in / 280 mm
1.37 in / 34.8 mm
6.65 in / 169 mm
Length
10.71 in / 272 mm
2.78 in / 70.7 mm
5 in / 127 mm
Weight
1100 g
28 g
470 g
Storage temperature
-13 to 158 °F
-25 to 70 °C
36 to 77 °F
2 to 25 °C
-13 to 158 °F
-25 to 70 °C
Relative humidity
(storage)
< 93%
< 93%
< 93%
Battery voltage/type
N/A
Not replaceable
N/A
Interfaces
N/A
IR port; Code key socket LED (red, green, blue)
Input voltage
N/A
N/A
+ 12 V DC
Input current
N/A
N/A
2.0 A
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......