Supplement for Observed Test Sequence
189
Operator:
6
Touch
Glucose Test
.
Perform the patient test as usual. Once the test is
completed, the observer has to complete the next steps.
7
Hand the meter back to the observer.
Observer Login
I-301: Observer
Alan
Smithee
, please log in to
proceed. Otherwise the
OTS test will not be
complete.
OTS
Valid Glucose Result
No
Yes
OTS
Pass
Fail
12:48 pm
09/18/13
Patient Test - OTS
Range
150
Patient 123456789
Date 09/18/13 12:48 pm
mg/dL
09/18/13
12:48 pm
Observer:
8
Touch
to log in again.
9
After you enter your password, touch
to
proceed with the assessment
.
10
Assess the test result validity by touching
Yes
or
No
.
11
Assess the operator’s performance by touching
Pass
or
Fail
.
12
Touch
, if you want to add a comment.
13
Touch
to return to the
Main Menu
screen.
The OTS information is saved together with the test result.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......