Supplement for Observed Test Sequence
188
Using the OTS function
A request for an Observed Test Sequence comes from the
DMS.
Patient Test Menu
Observer Login
Glucose Test
Other Test
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
A-O
0
Observer ID
12:48 pm
09/18/13
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
09/18/13
Observer:
1
Touch
Patient Test
.
In the
Patient Test Menu
, the
Glucose Test
button is
grayed out (disabled) until the observer has logged in.
2
Touch
Observer Login.
3
Wait until the
Observer ID
screen is displayed.
Enter your operator ID or scan your operator ID
barcode. Enter the (optional) password in the
Observer Password screen, and touch
to log
in.
The
Patient Test Menu
is displayed again. The
Glucose
Test
button is now available.
5
Hand the meter to the operator who can now
perform the patient test under supervision.
Other Test
Patient Test Menu
Observer Login
Glucose Test
12:48 pm
09/18/13
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......