Patient Glucose Testing
39
For emergencies, Short Turn-Around Time tests (STAT)
tests can be configured in the meter. This option allows
the meter to perform a limited number of patient glucose
tests, if cir cumstances require, despite the meter being in
QC Lock out or Download Lockout (see page 63).
In Measurement mode only: If you are performing a
test (patient, control, proficiency or linearity), the
meter will power off after 10 minutes of inactivity (no
screen touches), independently of configured
automatic power off time. If a result is already present,
the meter will emit three warning beeps every minute
after 5 minutes of inactivity and will save the result
before powering off after 10 minutes of activity.
n
Observe the applicable regulations and directives
for hygiene and safety when collecting blood
samples.
n
Observe the applicable regulations and directives
for disposing of potentially infectious samples and
materials.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......